Skip to content
Solution: TSS and CCTV monitoring with automatic intervention, for construction sites under Earth Control Measures.See how it works
Ebro · Data logger

EBI 12 temperature and pressure loggers

The Ebro EBI 12 temperature and pressure loggers are data loggers for recording temperature and pressure in sterilisation validation, in places such as steam sterilisers, autoclaves, endoscope washers and washer disinfectors. They record with a class A PT1000 temperature sensor and a piezoresistive pressure sensor covering 1 to 4,000 mbar. The versions differ in probe or needle, which suits them to different chambers, drains and washers.

Temperature range0 °C … +150 °Cvaries by version
Pressure range1 mbar … 4,000 mbarvaries by version
Temperature accuracy± 0.05 … ± 0.1 °Cvaries by version
Data memory2x 33,440 · 4x 20,064 measurementsvaries by version
IP ratingIP68
EBI 12-TP222: ebro logger in stainless steel housing with a long coiled temperature probe

Highlights

01
Temperature and pressure logger

One logger that records temperature and pressure, with a choice of probe type, length and connection across the versions.

02
PT1000 and piezoresistive sensors

Temperature uses a class A PT1000 sensor at 0.01 °C resolution. Pressure uses a piezoresistive sensor at 1 mbar resolution.

03
Measurement interval and real-time data

The measurement interval runs from 250 ms to 24 hours. Real-time data transmission is supported.

04
Factory calibration certificate included

The certificate covers temperature at 0 °C, +60 °C and +134 °C, and pressure at 100 mbar and 3,100 mbar.

Parameters

T

Temperature

Range0 °C … +150 °C
Accuracy± 0.05 … ± 0.1 °C
Resolution0.01 °C
SensorPT1000, class A
p

Pressure

Range1 mbar … 4,000 mbar
Accuracyfrom ± 10 mbar
Resolution1 mbar
SensorPiezoresistive pressure sensor

Configurations

Every version has the same class A PT1000 temperature sensor, piezoresistive pressure sensor, 1 to 4,000 mbar pressure range, IP68 rating and 250 ms to 24 hours measurement interval, so the choice is between probe or needle type, connection and temperature range.

1340-6653

EBI 12-TP222

Bendable metal probe, 1.50 x 500 mm, Luer-Lock connection. Range 0 °C … +150 °C. For autoclaves and endoscope washers.

1340-6657

EBI 12-TP226

Bendable metal probe, 1.50 x 250 mm. Range 0 °C … +150 °C. For the steriliser drain.

1340-6655-EX

EBI 12-TP231-EX

Needle, 1.95 x 40 mm, Luer-Lock connection. Range 0 °C … +85 °C. ATEX approved for explosive areas.

1340-6652

EBI 12-TP234

Needle, 1.95 x 40 mm, hose connection. Range 0 °C … +150 °C. For washer disinfectors.

1340-6662

EBI 12-TP422

Three bendable metal probes, 1.50 x 500 mm, Luer-Lock connection. Range 0 °C … +150 °C.

1340-6655

EBI 12-TP231

Needle, 40 mm. Range 0 °C … +150 °C. For steam sterilisers and washer disinfectors.

1340-6658

EBI 12-TP237

Needle, 1.50 x 20 mm, hose connection. Range 0 °C … +140 °C. For routine and batch control.

Specifications

Operating temperatureEBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP4220 °C … +150 °C
Measurement range: temperatureEBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP2310 °C … +150 °CEBI 12-TP2370 °C … +140 °C
Measurement range: pressureEBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP231EBI 12-TP2371 … 4,000 mbar
Accuracy: temperatureEBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP422EBI 12-TP231± 0.1 °C (0 °C … +120 °C) ± 0.05 °C (+120 °C … +140 °C) ± 0.1 °C (+140 °C … +150 °C)
Accuracy: pressureEBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP422± 15 mbar (1 mbar … 49 mbar) ± 10 mbar (50 mbar … 150 mbar) ± 15 mbar (151 mbar … 2,049 mbar) ± 10 mbar (2,050 mbar … 2,250 mbar) ± 15 mbar (2,241 mbar … 2,999 mbar) ± 10 mbar (3,000 mbar … 3,250 mbar) ± 15 mbar (3,251 mbar … 3,499 mbar) ± 20 mbar (3,500 mbar … 4,000 mbar)EBI 12-TP231±15 mbar (1 mbar … 50 mbar) ±10 mbar (50 mbar … 150 mbar) ±10 mbar (2,050 mbar … 2,250 mbar) ±10 mbar (3,000 mbar … 3,250 mbar) ± 15 mbar (3,250 mbar … 3,500 mbar) ± 0.5 % FS (3,500 mbar … 4,000 mbar)EBI 12-TP237±20 mbar
Resolution: temperature0.01 °C
Resolution: pressureEBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP422EBI 12-TP2311 mbar
Data memoryEBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP231EBI 12-TP2372x 33,440 measurementsEBI 12-TP4224x 20,064 measurements
Sensor: temperaturePT1000, class A
Sensor: pressurePiezoresistive pressure sensor
Measurement interval250 ms … 24 hours
Real-time data transmissionYes
Storage temperature-20 °C … +70 °C
Measurement modesEBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP422Endless measurement; Measure upon starttime; Start immediately until end of memory; Start- / Stop measurement
BatteryLithium, 3.6 V, User-replaceable
Dimensions (Ø x H)EBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP42248 mm x 24 mmEBI 12-TP231(Ø x H) 48 mm x 32 mm - Needle length 40 mmEBI 12-TP237ø 48 mm x 32 mm - Needle length 20 mm
WeightEBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP422app. 90 gEBI 12-TP231EBI 12-TP237approx. 120 g
Housing materialEBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP422316 L stainless steel / PEEK
IP ratingIP68
Factory calibration certificate0 °C, +60 °C and +134 °C 100 mbar and 3,100 mbar (at +25 °C) 3,100 mbar (at +134 °C)
EN 285EBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP422Sterilization, steam sterilizers, medical equipment
EN 13060EBI 12-TP222EBI 12-TP226EBI 12-TP231-EXEBI 12-TP234EBI 12-TP422Small steam sterilizers
ISO 17665Sterilization of health care products — Moist heat
ISO 15883EBI 12-TP222EBI 12-TP231-EXEBI 12-TP234EBI 12-TP422EBI 12-TP231EBI 12-TP237Washer Disinfectors
2014/53/EUEBI 12-TP222EBI 12-TP226EBI 12-TP234EBI 12-TP422EBI 12-TP237Radio Equipment Directive - RED
Type descriptionEBI 12-TP222L = 500 mm, Luer-Lock-connectionEBI 12-TP226L = 250 mmEBI 12-TP231-EXL = 40 mm, Luer-Lock-connectionEBI 12-TP234L = 40 mm, Hose connectionEBI 12-TP422L = 500 mm, Luer-Lock connection
Temperature sensor diameterEBI 12-TP222EBI 12-TP226EBI 12-TP237Ø1.50 mmEBI 12-TP231-EXEBI 12-TP234Ø1.95 mm
ATEX approvedYes
Permissible ambient conditions in the hazardous area0 °C … +85 °C; 100 Pa … 400 kPa
ATEX-TypeII1G Ex ia IIC T4 Ga
2014/34/EUATEX Directive
(EC) 1935/2004Materials and articles which come into contact with foodstuffs

Standards

EN 285

Sterilization, steam sterilizers, medical equipment

EN 13060

Small steam sterilizers

ISO 17665

Sterilization of health care products — Moist heat

ISO 15883

Washer Disinfectors

Tags and applications

Measured variable
TemperaturePressure
Form factor
Data logger
Brand
Ebro

Downloads

Common questions

How many measurements can the EBI 12 temperature and pressure loggers store?

The EBI 12-TP222, EBI 12-TP226, EBI 12-TP231-EX, EBI 12-TP234, EBI 12-TP231 and EBI 12-TP237 store 2x 33,440 measurements. The EBI 12-TP422 stores 4x 20,064 measurements.

What measurement interval can I set on the EBI 12 temperature and pressure loggers?

The measurement interval runs from 250 ms to 24 hours.

What pressure range do the EBI 12 temperature and pressure loggers measure?

The pressure range is 1 to 4,000 mbar with a resolution of 1 mbar. The pressure sensor is piezoresistive.

Is there an ATEX version of the EBI 12 temperature and pressure loggers?

Yes. The EBI 12-TP231-EX is ATEX approved as II1G Ex ia IIC T4 Ga under 2014/34/EU. Its permissible ambient conditions in the hazardous area are 0 °C to +85 °C and 100 Pa to 400 kPa.

What does the factory calibration certificate cover?

Temperature is certified at 0 °C, +60 °C and +134 °C. Pressure is certified at 100 mbar and 3,100 mbar at +25 °C, and at 3,100 mbar at +134 °C.

Can the battery in the EBI 12 temperature and pressure loggers be replaced?

Yes. They use a 3.6 V lithium battery that the user can replace. They are rated IP68.

Tell us what you need to measure.

Send the parameters and ranges you need, and we will confirm the right configuration and price. Your enquiry will name this product.