Skip to content
Solution: TSS and CCTV monitoring with automatic intervention, for construction sites under Earth Control Measures.See how it works
Ebro · Data logger

EBI 12-TC230

The Ebro EBI 12-TC230 is a conductivity and temperature data logger for checking the final rinse water in washer disinfectors when medical devices are reprocessed, because the rinse water's conductivity shows how well the cleaning worked. It measures during the programme without interrupting it, from 1 to 2,000 µS/cm and from 0 °C to +125 °C. Ebro recommends using it with the AL 132 flow adapter.

Temperature range0 °C … +125 °C
Conductivity range1 µS/cm … 2,000 µS/cm
Temperature accuracy± 0.2 °C
Data memory2x 50,160 measurements
IP ratingIP68
stainless steel ebro logger with probe tube, side view

Highlights

01
Temperature and conductivity in one logger

One logger records temperature with a PT1000 class A sensor and conductivity with a conductivity electrode.

02
IP68 stainless steel housing

The housing is 316 L stainless steel and PEEK with IP68 protection, 48 mm in diameter and 130 mm high.

03
Memory and measurement interval

Data memory is 2x 50,160 measurements, and the measurement interval can be set from 1 s to 24 hours.

04
Four measurement modes

Modes are endless measurement, measure upon start time, start immediately until end of memory, and start/stop measurement.

Parameters

T

Temperature

Range0 °C … +125 °C
Accuracy± 0.2 °C
Resolution0.01 °C
SensorPT1000, class A
κ

Conductivity

Range1 µS/cm … 2,000 µS/cm
Accuracyfrom ± 0.5 µS/cm
Resolution0.1 µS/cm
SensorConductivity electrode

Accuracy across the range

Accuracy over each temperature range. The bar shows where the range sits on the scale.

+1 … +20 µS/cmMost accurate± 0.5 µS/cm
+20 … +100 µS/cm± 1.0 µS/cm
+100 … +1500 µS/cm± 3.0 % from the measurement
+1500 … +2000 µS/cm± 5 % from the full scale
+1 µS/cm+500.75+1000.5+1500.25+2000 µS/cm

Specifications

Part number1340-6667
Operating temperature0 °C … +125 °C
Measuring range: temperature0 °C … +125 °C
Measuring range: conductivity1 µS/cm … 2,000 µS/cm
Accuracy: temperature± 0.2 °C
Accuracy: conductivity± 0.5 µS/cm (1 µS/cm … 20 µS/cm) ± 1.0 µS/cm (20.1 µS/cm … 100 µS/cm) ± 3.0 % of reading (100.1 µS/cm … 1,500 µS/cm) ± 5 % of full scale (rest of range)
Resolution: temperature0.01 °C
Resolution: conductivity0.1 µS/cm
Data memory2x 50,160 measurements
Sensor: temperaturePT1000, class A
Sensor: conductivityConductivity electrode
Measurement interval1s … 24 hours
Real-time data transmissionNo
Storage temperature-20 °C … +70 °C
Measurement modesEndless measurement; Measure upon starttime; Start immediately until end of memory; Start- / Stop measurement
BatteryLithium, 3.6 V, User-replaceable
Dimensions (Ø x H)48 mm x 130 mm
WeightApp. 150 g
Housing material316 L stainless steel / PEEK
Protection classIP68
Factory calibration certificate0 °C, +60 °C and +121 °C; 5.0 µS/cm, 100.0 µS/cm and 1,413.0 µS/cm
Washer disinfectorsISO 15883
Radio Equipment Directive RED2014/53/EU
Electromagnetic compatibility (EMC)DIN EN 301 489

Standards

ISO 15883

Washer Disinfectors

Tags and applications

Measured variable
Temperature
Form factor
Data logger
Brand
Ebro

Downloads

Common questions

What temperature range and accuracy does the EBI 12-TC230 have?

It measures 0 °C to +125 °C with an accuracy of ± 0.2 °C and a resolution of 0.01 °C.

What is the conductivity measuring range and accuracy?

The range is 1 µS/cm to 2,000 µS/cm. Accuracy is ± 0.5 µS/cm from 1 to 20 µS/cm, ± 1.0 µS/cm from 20.1 to 100 µS/cm, ± 3.0 % of the measurement from 100.1 to 1,500 µS/cm and ± 5 % of full scale above that.

Does the EBI 12-TC230 transmit data in real time?

No. The datasheet lists real-time data transmission as not available. Readings are stored in the logger's memory.

Which standard is the EBI 12-TC230 listed for?

The datasheet lists ISO 15883, the standard for washer disinfectors.

Can the battery be replaced by the user?

Yes. The battery is a 3.6 V lithium cell and is user-replaceable.

What points does the factory calibration certificate cover?

Temperature is certified at 0 °C, +60 °C and +121 °C. Conductivity is certified at 5.0 µS/cm, 100.0 µS/cm and 1,413.0 µS/cm.

Tell us what you need to measure.

Send the parameters and ranges you need, and we will confirm the right configuration and price. Your enquiry will name this product.