EBI 11 validation loggers
The Ebro EBI 11-T235, EBI 11-T236 and EBI 11-T240 are mini temperature data loggers for validating the reprocessing of medical devices, in places such as doctors' surgeries, the Sirona DAC Universal and Melag Careclave, cassette autoclaves and washer disinfectors. Each records with a Pt 1000 Class A sensor from -30 °C to +150 °C, on a 25 or 80 mm needle or, on the EBI 11-T240, a bendable 250 mm probe with a fast response time. The loggers are programmed and read out through an Ebro interface with Winlog.pro software.
Highlights
Stainless steel mini data logger with fixed sensor needle length for reproducible measurement, used with DAC Universal and Careclave.
Measures -30 °C to +150 °C at 0.01 °C resolution, with ±0.2 °C accuracy below 0 °C and ±0.1 °C from 0 °C upwards.
Measurement interval from 1 s to 24 hours, 15,000 measurements of memory, and endless, timed-start, run-to-memory-end and start/stop modes.
316 TI stainless steel housing rated IP68, with an M5 thread and a user-replaceable lithium battery.
Parameters
Temperature
Accuracy across the range
Accuracy over each temperature range. The bar shows where the range sits on the scale.
Configurations
Every version has the same Pt 1000 Class A sensor, measurement range, data memory and stainless steel housing, and the versions differ in needle length and probe.
EBI 11-T235
For the DAC Universal and Careclave, and the endoscope dummy; 25 mm needle.
EBI 11-T236
For the DAC Universal and Careclave; 80 mm needle.
EBI 11-T240
For cassette autoclaves and small sterilisers; bendable 250 mm probe, 1.50 mm diameter, fast response time.
Specifications
Standards
Small steam sterilizers
Sterilization of health care products — Moist heat
Washer Disinfectors
Tags and applications
- Measured variable
- Temperature
- Form factor
- Data logger
- Application
- Sterilisation validation
- Brand
- Ebro
Downloads
Common questions
What temperature range does the EBI 11 validation logger cover?
It measures from -30 °C to +150 °C with a Pt 1000 Class A sensor. Accuracy is ±0.2 °C from -30 °C to 0 °C and ±0.1 °C from 0 °C to +150 °C.
How many readings can the EBI 11 store?
The data memory holds 15,000 measurements. The measurement interval can be set from 1 s to 24 hours.
Can the EBI 11 be used to check endoscope washers?
Yes. The supplier states it can be used to check endoscope washers, and for process validation with DAC Universal and Careclave.
Which standards does the EBI 11 datasheet list?
It lists EN 13060 for small steam sterilizers, ISO 17665 for moist heat sterilisation of health care products, and ISO 15883 for washer disinfectors.
What battery does the EBI 11 use, and can I replace it?
It uses two BR1225A 3 V lithium batteries, and they are user-replaceable. The AL 113 battery exchange set has 6 batteries, 3 gaskets with lubricant and a screwdriver. The AL 113L set covers 10 battery changes, with 20 batteries, 10 O-rings with grease and a changing tool.
At which temperatures is the factory calibration certificate issued?
The factory calibration certificate covers 0 °C, +60 °C and +134 °C.
What do I need to program and read out the loggers?
An Ebro interface set with Winlog.pro software, connected by USB. The SI 1100 set has a 2-port IF 100 interface and the SI 1300 set a 4-port IF 300 interface. The SK 6100 starter kit contains an EBI 11-T210, an IF 300 interface and Winlog.pro, and takes further EBI 11 loggers.
Tell us what you need to measure.
Send the parameters and ranges you need, and we will confirm the right configuration and price. Your enquiry will name this product.