Skip to content
Solution: TSS and CCTV monitoring with automatic intervention, for construction sites under Earth Control Measures.See how it works
Ebro · Data logger

EBI 11 validation loggers

The Ebro EBI 11-T235, EBI 11-T236 and EBI 11-T240 are mini temperature data loggers for validating the reprocessing of medical devices, in places such as doctors' surgeries, the Sirona DAC Universal and Melag Careclave, cassette autoclaves and washer disinfectors. Each records with a Pt 1000 Class A sensor from -30 °C to +150 °C, on a 25 or 80 mm needle or, on the EBI 11-T240, a bendable 250 mm probe with a fast response time. The loggers are programmed and read out through an Ebro interface with Winlog.pro software.

Range-30 °C … +150 °C
Accuracy± 0.1 … ± 0.2 °C
Resolution0.01 °C
Data memory15,000 measurements
IP ratingIP68
EBI 11-T236: Stainless steel cylindrical logger with long thin probe rod

Highlights

01
Mini logger with fixed needle

Stainless steel mini data logger with fixed sensor needle length for reproducible measurement, used with DAC Universal and Careclave.

02
Pt 1000 Class A sensor

Measures -30 °C to +150 °C at 0.01 °C resolution, with ±0.2 °C accuracy below 0 °C and ±0.1 °C from 0 °C upwards.

03
Interval and start modes

Measurement interval from 1 s to 24 hours, 15,000 measurements of memory, and endless, timed-start, run-to-memory-end and start/stop modes.

04
IP68 housing with M5 thread

316 TI stainless steel housing rated IP68, with an M5 thread and a user-replaceable lithium battery.

Parameters

T

Temperature

Range-30 °C … +150 °C
Accuracy± 0.1 … ± 0.2 °C
Resolution0.01 °C
SensorPt 1000, Class A

Accuracy across the range

Accuracy over each temperature range. The bar shows where the range sits on the scale.

−30 … 0 °C±0.2 °C
0 … +150 °CMost accurate±0.1 °C
−30 °C+15+60+105+150 °C

Configurations

Every version has the same Pt 1000 Class A sensor, measurement range, data memory and stainless steel housing, and the versions differ in needle length and probe.

1340-6270

EBI 11-T235

For the DAC Universal and Careclave, and the endoscope dummy; 25 mm needle.

1340-6271

EBI 11-T236

For the DAC Universal and Careclave; 80 mm needle.

1340-6291

EBI 11-T240

For cassette autoclaves and small sterilisers; bendable 250 mm probe, 1.50 mm diameter, fast response time.

Specifications

Operating temperature-30 °C … +150 °C
Measurement range-30 °C … +150 °C
Accuracy±0.2 °C (-30 °C … 0 °C) ±0.1 °C (0 °C … +150 °C)
Resolution0.01 °C
Data memory15,000 measurements
SensorPt 1000, Class A
Measurement interval1 s … 24 hours
Storage temperature-20 °C … +70 °C
Measurement modeEndless measurement; Measure upon starttime; Start immediately until end of memory; Start- / Stop measurement
BatteryLithium, 2x BR1225A, 3V, user-replaceable
Dimensions (Ø x H)16.5 mm x 24 mm
Weightapp. 45 g
Housing material316 TI stainless steel
IP ratingIP68
Factory calibration certificate0 °C, +60 °C and +134 °C
Probe lengthEBI 11-T23525 mmEBI 11-T23680 mmEBI 11-T240250 mm
EN 13060Small steam sterilizers
ISO 17665Sterilization of health care products — Moist heat
ISO 15883Washer Disinfectors
FCM EN 1935/2004Food Material
2014/30/EUElectromagnetic compatibility of equipment

Standards

EN 13060

Small steam sterilizers

ISO 17665

Sterilization of health care products — Moist heat

ISO 15883

Washer Disinfectors

Tags and applications

Measured variable
Temperature
Form factor
Data logger
Brand
Ebro

Downloads

Common questions

What temperature range does the EBI 11 validation logger cover?

It measures from -30 °C to +150 °C with a Pt 1000 Class A sensor. Accuracy is ±0.2 °C from -30 °C to 0 °C and ±0.1 °C from 0 °C to +150 °C.

How many readings can the EBI 11 store?

The data memory holds 15,000 measurements. The measurement interval can be set from 1 s to 24 hours.

Can the EBI 11 be used to check endoscope washers?

Yes. The supplier states it can be used to check endoscope washers, and for process validation with DAC Universal and Careclave.

Which standards does the EBI 11 datasheet list?

It lists EN 13060 for small steam sterilizers, ISO 17665 for moist heat sterilisation of health care products, and ISO 15883 for washer disinfectors.

What battery does the EBI 11 use, and can I replace it?

It uses two BR1225A 3 V lithium batteries, and they are user-replaceable. The AL 113 battery exchange set has 6 batteries, 3 gaskets with lubricant and a screwdriver. The AL 113L set covers 10 battery changes, with 20 batteries, 10 O-rings with grease and a changing tool.

At which temperatures is the factory calibration certificate issued?

The factory calibration certificate covers 0 °C, +60 °C and +134 °C.

What do I need to program and read out the loggers?

An Ebro interface set with Winlog.pro software, connected by USB. The SI 1100 set has a 2-port IF 100 interface and the SI 1300 set a 4-port IF 300 interface. The SK 6100 starter kit contains an EBI 11-T210, an IF 300 interface and Winlog.pro, and takes further EBI 11 loggers.

Tell us what you need to measure.