EBI 12 steam steriliser loggers
The Ebro EBI 12 steam steriliser loggers are temperature data loggers for validating moist heat sterilisation in places such as steam sterilisers and, with the EBI 12-T421, EBI 12-T441 and EBI 12-T441-EX, washer disinfectors. They measure with a PT1000 class A sensor on bendable metal or flexible cable probes, depending on the version, and transmit readings in real time. Accuracy is ±0.05 °C between +120 °C and +140 °C.
Highlights
Every version has a PT1000 class A sensor and 0.01 °C resolution. The versions differ in range, probes and memory.
The factory certificate on every version covers 0 °C, +60 °C and +134 °C. Accuracy is ± 0.05 °C from +120 °C to +140 °C.
The 316 L stainless steel and PEEK housing is IP68 rated, measures 48 mm x 24 mm (Ø x H) and weighs approximately 90 g.
The measurement interval runs from 250 ms to 24 hours. Real-time data transmission is supported on every version.
Parameters
Temperature
Accuracy across the range
Accuracy over each temperature range. The bar shows where the range sits on the scale.
Configurations
Every version has the same PT1000 class A sensor, 0.01 °C resolution, IP68 housing in 316 L stainless steel and PEEK, and 250 ms to 24 hour measurement interval, and the versions differ in range, probes, memory and listed standards.
EBI 12-T421
Two bendable metal probes, 1.50 x 500 mm. Range -200 °C … +400 °C. Also for hot air sterilisers and baking.
EBI 12-T441
Two bendable metal probes, 1.50 x 500 mm. Range -200 °C … +200 °C.
EBI 12-T470
Two flexible cable probes, 1.20 x 600 mm. Range -20 °C … +150 °C.
EBI 12-T471
Two flexible cable probes, 1.20 x 1,200 mm. Range -20 °C … +150 °C.
EBI 12-T490
Two flexible cable probes, 1.20 x 600 mm. Range -20 °C … +150 °C.
EBI 12-T671
Four flexible cable probes, 1.20 x 1,200 mm. Range -20 °C … +150 °C.
EBI 12-T690
Four flexible cable probes, 1.20 x 600 mm. Range -20 °C … +150 °C.
EBI 12-T441-EX
Two bendable metal probes, 1.50 x 500 mm. Range -200 °C … +200 °C. ATEX approved for explosive areas.
Specifications
Standards
Sterilization of health care products — Moist heat
Washer Disinfectors
Tags and applications
- Measured variable
- Temperature
- Form factor
- Data logger
- Application
- Sterilisation validation
- Brand
- Ebro
Downloads
Datasheet, EBI 12-T421 (1340-6630)
Ebro · PDF · 236 KB
Datasheet, EBI 12-T441 (1340-6629)
Ebro · PDF · 214 KB
Datasheet, EBI 12-T470 (1340-6640)
Ebro · PDF · 235 KB
Datasheet, EBI 12-T471 (1340-6644)
Ebro · PDF · 235 KB
Datasheet, EBI 12-T490 (1340-6634)
Ebro · PDF · 228 KB
Datasheet, EBI 12-T671 (1340-6645)
Ebro · PDF · 220 KB
Datasheet, EBI 12-T690 (1340-6635)
Ebro · PDF · 220 KB
Datasheet, EBI 12-T441-EX (1340-6629-EX)
Ebro · PDF · 222 KB
Common questions
What temperature range do the EBI 12 steam steriliser loggers cover?
The EBI 12-T421 measures -200 °C to +400 °C. The EBI 12-T441 and EBI 12-T441-EX measure -200 °C to +200 °C. The EBI 12-T470, EBI 12-T471, EBI 12-T490, EBI 12-T671 and EBI 12-T690 measure -20 °C to +150 °C.
How accurate are the EBI 12 steam steriliser loggers at 134 °C?
Accuracy is ± 0.05 °C between +120 °C and +140 °C on every version. Resolution is 0.01 °C. The sensor is a PT1000, class A. The factory calibration certificate includes a point at +134 °C on every version.
How many measurements can the EBI 12 steam steriliser loggers store?
The EBI 12-T421, EBI 12-T441, EBI 12-T441-EX, EBI 12-T470, EBI 12-T471 and EBI 12-T490 store 2x 50,160 measurements. The EBI 12-T671 and EBI 12-T690 store 4x 25,080 measurements.
Can I replace the battery in the EBI 12 steam steriliser loggers myself?
Yes. The battery is a 3.6 V lithium cell and it is user-replaceable.
Is there an ATEX version of the EBI 12 steam steriliser loggers for explosive areas?
Yes. The EBI 12-T441-EX is ATEX approved under Directive 2014/34/EU, marked II1G Ex ia IIC T4 Ga. It has two bendable metal probes, 1.50 x 500 mm, and a range of -200 °C to +200 °C.
What measurement interval and start modes do the EBI 12 steam steriliser loggers offer?
The measurement interval is adjustable from 250 ms to 24 hours. The modes are endless measurement, measure upon start time, start immediately until end of memory, and start / stop measurement. They also support real-time data transmission.
Tell us what you need to measure.