Skip to content
Solution: TSS and CCTV monitoring with automatic intervention, for construction sites under Earth Control Measures.See how it works
Ebro · Validation data logger

EBI 12 steam steriliser loggers

The Ebro EBI 12 steam steriliser loggers are temperature data loggers for validating moist heat sterilisation in places such as steam sterilisers and, with the EBI 12-T421, EBI 12-T441 and EBI 12-T441-EX, washer disinfectors. They measure with a PT1000 class A sensor on bendable metal or flexible cable probes, depending on the version, and transmit readings in real time. Accuracy is ±0.05 °C between +120 °C and +140 °C.

Range-200 °C … +400 °Cvaries by version
Accuracy± 0.05 … ± 1.5 °Cvaries by version
Resolution0.01 °C
Data memory2x 50,160 · 4x 25,080 measurementsvaries by version
IP ratingIP68
EBI 12-T421: ebro logger with two coiled rigid stainless steel probes

Highlights

01
One logger, several probe sets

Every version has a PT1000 class A sensor and 0.01 °C resolution. The versions differ in range, probes and memory.

02
Factory calibration at +134 °C

The factory certificate on every version covers 0 °C, +60 °C and +134 °C. Accuracy is ± 0.05 °C from +120 °C to +140 °C.

03
IP68 stainless steel housing

The 316 L stainless steel and PEEK housing is IP68 rated, measures 48 mm x 24 mm (Ø x H) and weighs approximately 90 g.

04
Interval range and real-time transmission

The measurement interval runs from 250 ms to 24 hours. Real-time data transmission is supported on every version.

Parameters

T

Temperature

Range-200 °C … +400 °C
Accuracy± 0.05 … ± 1.5 °C
Resolution0.01 °C
SensorPT1000, class A

Accuracy across the range

Accuracy over each temperature range. The bar shows where the range sits on the scale.

−200 … −90 °C± 1.5 °C
−90 … −40 °C± 0.5 °C
−40 … 0 °C± 0.2 °C
0 … +120 °C± 0.1 °C
+120 … +140 °CMost accurate± 0.05 °C
+140 … +150 °C± 0.1 °C
+150 … +250 °C± 0.5 °C
+250 … +400 °C± 0.8 °C
−200 °C−50+100+250+400 °C

Configurations

Every version has the same PT1000 class A sensor, 0.01 °C resolution, IP68 housing in 316 L stainless steel and PEEK, and 250 ms to 24 hour measurement interval, and the versions differ in range, probes, memory and listed standards.

1340-6630

EBI 12-T421

Two bendable metal probes, 1.50 x 500 mm. Range -200 °C … +400 °C. Also for hot air sterilisers and baking.

1340-6629

EBI 12-T441

Two bendable metal probes, 1.50 x 500 mm. Range -200 °C … +200 °C.

1340-6640

EBI 12-T470

Two flexible cable probes, 1.20 x 600 mm. Range -20 °C … +150 °C.

1340-6644

EBI 12-T471

Two flexible cable probes, 1.20 x 1,200 mm. Range -20 °C … +150 °C.

1340-6634

EBI 12-T490

Two flexible cable probes, 1.20 x 600 mm. Range -20 °C … +150 °C.

1340-6645

EBI 12-T671

Four flexible cable probes, 1.20 x 1,200 mm. Range -20 °C … +150 °C.

1340-6635

EBI 12-T690

Four flexible cable probes, 1.20 x 600 mm. Range -20 °C … +150 °C.

1340-6629-EX

EBI 12-T441-EX

Two bendable metal probes, 1.50 x 500 mm. Range -200 °C … +200 °C. ATEX approved for explosive areas.

Specifications

Operating temperatureEBI 12-T421EBI 12-T441EBI 12-T441-EX-55 °C … +150 °CEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T690-20 °C … +150 °C
Measurement rangeEBI 12-T421-200 °C … +400 °CEBI 12-T441EBI 12-T441-EX-200 °C … +200 °CEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T690-20 °C … +150 °C
AccuracyEBI 12-T421± 1.5 °C (-200 °C … -90 °C) ± 0.5 °C (-90 °C … -40 °C) ± 0.2 °C (-40 °C … 0 °C) ± 0.1 °C (0 °C … +120 °C) ± 0.05 °C (+120 °C … +140 °C) ± 0.1 °C (+140 °C … +150 °C) ± 0.5 °C (+150 °C … +250 °C) ± 0.8 °C (+250 °C … +400 °C)EBI 12-T441EBI 12-T441-EX± 1.5 °C (-200 °C … -90 °C) ± 0.5 °C (-90 °C … -40 °C) ± 0.2 °C (-40 °C … 0 °C) ± 0.1 °C (0 °C … +120 °C) ± 0.05 °C (+120 °C … +140 °C) ± 0.1 °C (+140 °C … +150 °C) ± 0.5 °C (+150 °C … +200 °C)EBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T690± 0.2 °C (-20 °C … 0 °C) ± 0.1 °C (0 °C … +120 °C) ± 0.05 °C (+120 °C … +140 °C) ± 0.1 °C (+140 °C … +150 °C)
Resolution0.01 °C
Data memoryEBI 12-T421EBI 12-T441EBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T441-EX2x 50,160 measurementsEBI 12-T671EBI 12-T6904x 25,080 measurements
SensorPT1000, class A
Measurement interval250 ms … 24 hours
Real-time data transmissionYes
Storage temperature-20 °C … +70 °C
Measurement modesEndless measurement; Measure upon start time; Start immediately until end of memory; Start- / Stop measurement
BatteryLithium, 3.6 V, User-replaceable
Dimensions (Ø x H)48 mm x 24 mm
WeightApp. 90 g
Housing material316 L stainless steel / PEEK
IP ratingIP68
Factory calibration certificateEBI 12-T421EBI 12-T441EBI 12-T441-EX-20 °C, 0 °C, +60 °C and +134 °CEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T6900 °C, +60 °C and +134 °C
Food contact materials regulationEBI 12-T421EBI 12-T441EBI 12-T441-EX(EC) 1935/2004 – Materials and articles which come into contact with food
Sterilisation standardISO 17665 – Sterilization of health care products — Moist heat
Washer disinfector standardEBI 12-T421EBI 12-T441EBI 12-T441-EXISO 15883 – Washer Disinfectors
Radio equipment directiveEBI 12-T421EBI 12-T441EBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T6902014/53/EU – Radio Equipment Directive RED
ATEX approvalEBI 12-T441-EXII1G Ex ia IIC T4 Ga, Directive 2014/34/EU
Probe diameterEBI 12-T421EBI 12-T441EBI 12-T441-EX1.50 mmEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T6901.20 mm
Probe lengthEBI 12-T421EBI 12-T441EBI 12-T441-EX500 mmEBI 12-T470EBI 12-T490EBI 12-T690600 mmEBI 12-T471EBI 12-T6711,200 mm
Steam steriliser standardEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T690EN 285 – Sterilization, steam sterilizers, medical equipment
Small steam steriliser standardEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T690EN 13060 – small steam sterilizers

Standards

ISO 17665

Sterilization of health care products — Moist heat

ISO 15883

Washer Disinfectors

Tags and applications

Measured variable
Temperature
Form factor
Data logger
Brand
Ebro

Downloads

Common questions

What temperature range do the EBI 12 steam steriliser loggers cover?

The EBI 12-T421 measures -200 °C to +400 °C. The EBI 12-T441 and EBI 12-T441-EX measure -200 °C to +200 °C. The EBI 12-T470, EBI 12-T471, EBI 12-T490, EBI 12-T671 and EBI 12-T690 measure -20 °C to +150 °C.

How accurate are the EBI 12 steam steriliser loggers at 134 °C?

Accuracy is ± 0.05 °C between +120 °C and +140 °C on every version. Resolution is 0.01 °C. The sensor is a PT1000, class A. The factory calibration certificate includes a point at +134 °C on every version.

How many measurements can the EBI 12 steam steriliser loggers store?

The EBI 12-T421, EBI 12-T441, EBI 12-T441-EX, EBI 12-T470, EBI 12-T471 and EBI 12-T490 store 2x 50,160 measurements. The EBI 12-T671 and EBI 12-T690 store 4x 25,080 measurements.

Can I replace the battery in the EBI 12 steam steriliser loggers myself?

Yes. The battery is a 3.6 V lithium cell and it is user-replaceable.

Is there an ATEX version of the EBI 12 steam steriliser loggers for explosive areas?

Yes. The EBI 12-T441-EX is ATEX approved under Directive 2014/34/EU, marked II1G Ex ia IIC T4 Ga. It has two bendable metal probes, 1.50 x 500 mm, and a range of -200 °C to +200 °C.

What measurement interval and start modes do the EBI 12 steam steriliser loggers offer?

The measurement interval is adjustable from 250 ms to 24 hours. The modes are endless measurement, measure upon start time, start immediately until end of memory, and start / stop measurement. They also support real-time data transmission.

Tell us what you need to measure.

Send the parameters and ranges you need, and we will confirm the right configuration and price. Your enquiry will name this product.