Skip to content
Solution: TSS and CCTV monitoring with automatic intervention, for construction sites under Earth Control Measures.See how it works
Ebro · Data logger

EBI 12 single-channel loggers

The Ebro EBI 12 single-channel loggers are PT1000 temperature data loggers for sterilisation validation, in places such as moist heat steriliser cycles and washer disinfectors. Each logger records temperature with a class A sensor in an IP68 housing of 316 L stainless steel and PEEK, and keeps the readings in its own memory. The versions differ in probe type, length and range, and include a radial probe and an ATEX-approved logger for explosive areas.

Range-200 °C … +400 °Cvaries by version
Accuracy± 0.05 … ± 1.5 °Cvaries by version
Data memory100,320 · 27,840 measurementsvaries by version
Measurement interval250 ms … 24 · 1 s … 24 hoursvaries by version
SensorPT1000, class A
IP ratingIP68
EBI 12-T220-EX: Stainless steel ebro logger with flexible wire probe, shown with an Ex explosion protection mark

Highlights

01
Single-channel PT1000 temperature logger

Every version records temperature on one channel with a PT1000 class A sensor at 0.01 °C resolution. Probe, range and memory differ by version.

02
IP68 stainless steel housing

The housing is 316 L stainless steel and PEEK, rated IP68, and measures 48 mm in diameter by 24 mm in height.

03
Sterilisation and washer standards

The standards listed are ISO 17665 for moist heat sterilisation and ISO 15883 for washer disinfectors.

04
Measurement modes and replaceable battery

Modes are endless measurement, start at a set time, start until memory is full, and start/stop. The 3.6 V lithium battery is user-replaceable.

Parameters

T

Temperature

Range-200 °C … +400 °C
Accuracy± 0.05 … ± 1.5 °C
Resolution0.01 °C
SensorPT1000, class A
Channels1

Accuracy across the range

Accuracy over each temperature range. The bar shows where the range sits on the scale.

EBI 12-T220, EBI 12-T220-EX, EBI 12-T240, EBI 12-T241
−200 … −90 °C± 1.5 °C
−90 … −40 °C± 0.5 °C
−40 … 0 °C± 0.2 °C
0 … +120 °C± 0.1 °C
+120 … +140 °CMost accurate± 0.05 °C
+140 … +150 °C± 0.1 °C
+150 … +200 °C± 0.5 °C
−200 °C−50+100+250+400 °C

Configurations

Every version is a single-channel PT1000 class A temperature logger with 0.01 °C resolution, an IP68 housing in 316 L stainless steel and PEEK, and a user-replaceable 3.6 V lithium battery.

1340-6604

EBI 12-T220

Bendable metal probe, 250 mm. Range -200 °C … +200 °C. For the steriliser outlet.

1340-6604-EX

EBI 12-T220-EX

Bendable metal probe, 250 mm. Range -200 °C … +200 °C. ATEX approved for explosive areas, -40 °C … +85 °C there.

1340-6605

EBI 12-T221

Bendable metal probe, 500 mm. Range -200 °C … +400 °C. For hot air sterilisers, in the TIB2 protection box.

1340-6610

EBI 12-T222

Rigid radial probe, 1.50 x 40 mm, M5 thread. Range -55 °C … +140 °C. Placed inside the machine.

1340-6611

EBI 12-T240

Bendable metal probe, 250 mm. Range -200 °C … +200 °C. Goes into hollow parts, for measuring on the product.

1340-6612

EBI 12-T241

Bendable metal probe, 500 mm. Range -200 °C … +200 °C. Goes into hollow parts, for measuring on the product.

Specifications

Operating temperatureEBI 12-T220EBI 12-T220-EXEBI 12-T221EBI 12-T240EBI 12-T241-55 °C … +150 °CEBI 12-T222-55 °C … +140 °C
Measurement rangeEBI 12-T220EBI 12-T220-EXEBI 12-T240EBI 12-T241-200 °C … +200 °CEBI 12-T221-200 °C … +400 °CEBI 12-T222-55 °C … +140 °C
AccuracyEBI 12-T220EBI 12-T220-EXEBI 12-T240EBI 12-T241± 1.5 °C (-200 °C … -90 °C) ± 0.5 °C (-90 °C … -40 °C) ± 0.2 °C (-40 °C … 0 °C) ± 0.1 °C (0 °C … +120 °C) ± 0.05 °C (+120 °C … +140 °C) ± 0.1 °C (+140 °C … +150 °C) ± 0.5 °C (+150 °C … +200 °C)EBI 12-T221± 1.5 °C (-200 °C … -90 °C) ± 0.5 °C (-90 °C … -40 °C) ± 0.2 °C (-40 °C … 0 °C) ± 0.1 °C (0 °C … +120 °C) ± 0.05 °C (+120 °C … +140 °C) ± 0.1 °C (+140 °C … +150 °C) ± 0.5 °C (+150 °C … +250 °C) ± 0.8 °C (+250 °C … +400 °C)EBI 12-T222± 0.5 °C (-55 °C … -40 °C) ± 0.2 °C (-40 °C … 0 °C) ± 0.1 °C (0 °C … +140 °C)
Resolution0.01 °C
Data memoryEBI 12-T220EBI 12-T220-EXEBI 12-T221EBI 12-T240EBI 12-T241100,320 measurementsEBI 12-T22227,840 measurements
SensorPT1000, class A
Measurement intervalEBI 12-T220EBI 12-T220-EXEBI 12-T221EBI 12-T240EBI 12-T241250 ms … 24 hoursEBI 12-T2221 s … 24 hours
Real-time data transmissionEBI 12-T220EBI 12-T220-EXEBI 12-T221EBI 12-T240EBI 12-T241YesEBI 12-T222No
Storage temperature-20 °C … +70 °C
Measurement modesEndless measurement; Measure upon starttime; Start immediately until end of memory; Start- / Stop measurement
BatteryLithium, 3.6 V, User-replaceable
Dimensions (Ø x H)48 mm x 24 mm
Weightapp. 90 g
Housing material316 L stainless steel / PEEK
IP ratingIP68
Factory calibration certificateEBI 12-T220EBI 12-T220-EXEBI 12-T222EBI 12-T240EBI 12-T241-20 °C, 0 °C, +60 °C and +134 °CEBI 12-T221-20 °C, 0 °C, +60 °C, +134 °C and +250°C
Food contact materials regulationEBI 12-T220EBI 12-T220-EXEBI 12-T221EBI 12-T240EBI 12-T241(EC) 1935/2004
Sterilisation standardISO 17665 (Sterilization of health care products — Moist heat)
Washer disinfector standardISO 15883
Radio equipment directiveEBI 12-T220EBI 12-T221EBI 12-T222EBI 12-T240EBI 12-T2412014/53/EU (RED)
Measuring channels1
Probe diameter1.50 mm
Probe lengthEBI 12-T220EBI 12-T220-EXEBI 12-T240L = 250 mmEBI 12-T221EBI 12-T241L = 500 mmEBI 12-T222L = 40 mm
ATEX approvedYes
Permissible ambient conditions in the hazardous area-40 °C … +85 °C; 100 Pa … 400 kPa
ATEX-TypeII1G Ex ia IIC T4 Ga
ATEX Directive2014/34/EU
Steam steriliser standardEN 285 (Sterilization, steam sterilizers, medical equipment)
Small steam steriliser standardEN 13060 (small steam sterilizers)
VersionRadial version
MountingM5 thread for mounting on the sensor

Standards

ISO 17665

Sterilization of health care products — Moist heat

ISO 15883

Washer Disinfectors

Tags and applications

Measured variable
Temperature
Form factor
Data logger
Brand
Ebro

Downloads

Common questions

What temperature range do the EBI 12 single-channel loggers measure?

The EBI 12-T220, T220-EX, T240 and T241 measure -200 °C … +200 °C. The EBI 12-T221 measures -200 °C … +400 °C. The EBI 12-T222 measures -55 °C … +140 °C.

What measurement interval can I set on the EBI 12 single-channel loggers?

The EBI 12-T220, T220-EX, T221, T240 and T241 log at intervals from 250 ms to 24 hours. The EBI 12-T222 logs at intervals from 1 s to 24 hours.

Can I replace the battery in the EBI 12 single-channel loggers myself?

Yes. The battery is a 3.6 V lithium cell that the user can replace.

Is there an ATEX version for explosive areas?

Yes. The EBI 12-T220-EX is ATEX approved as II1G Ex ia IIC T4 Ga under Directive 2014/34/EU. The permissible ambient conditions in the hazardous area are -40 °C … +85 °C and 100 Pa … 400 kPa.

At which temperatures is the factory calibration certificate issued?

The certificate covers -20 °C, 0 °C, +60 °C and +134 °C. The EBI 12-T221 certificate also covers +250 °C.

What is the IP rating of the EBI 12 single-channel loggers and what is the housing made of?

The rating is IP68. The housing is 316 L stainless steel and PEEK. The logger measures 48 mm in diameter by 24 mm in height.

Tell us what you need to measure.

Send the parameters and ranges you need, and we will confirm the right configuration and price. Your enquiry will name this product.