TB350
The Lovibond TB350 is a turbidity meter for water testing in places such as drinking water treatment and distribution, swimming pools, and wastewater and trade effluent. It measures turbidity by nephelometric 90° detection on a 10 mL sample, across 0.01 to 4,000 NTU. Infrared versions use an IR LED (860 nm) to ISO 7027, white LED versions are EPA approved as an alternative to US EPA 180.1, and each light source comes as a portable battery version or a benchtop version with a power supply.
Highlights
Choose an IR or white LED light source, and a battery pack or a power supply. The measuring range is the same on all.
Resolution is 0.01 NTU up to 9.99 NTU, 0.1 NTU up to 99.9 NTU and 1 NTU from 100 to 4000 NTU.
Accuracy is ± 1.8 % of reading plus straylight. Repeatability is < 1 % or ± 0.01 NTU. Stray light is < 0.014 NTU.
It stores 250 measurements and all calibrations, verifications and factory-restore actions. Data transfer format is .csv; interfaces are USB-A and USB-C.
Parameters
Turbidity
The meter can display results in NTU, FNU, degrees, mg/L Kaolin or mg/L PSL.
Configurations
Every version has the same measuring range, accuracy and touchscreen, so you are choosing only the light source and whether it runs on a battery pack or a power supply.
IR with rechargeable battery pack
IR LED (860 nm) light source to ISO 7027, portable, supplied with a Ni-MH battery pack.
IR with power supply
IR LED (860 nm) light source to ISO 7027, benchtop, supplied with a power supply.
WL with rechargeable battery pack
White LED light source, EPA approved (alternative to US EPA 180.1), Ni-MH battery pack and an extra 2000 NTU T-CAL® standard.
WL with power supply
White LED light source, EPA approved (alternative to US EPA 180.1), benchtop, supplied with a power supply and a 2000 NTU T-CAL® standard.
Consumables
These are the consumables this instrument uses. We confirm the part numbers for your instrument when we quote.
Calibration standards
Calibration standards, pH buffers and electrolyte solutions for calibrating the instrument.
Cuvettes & sample cells
Round cells, cuvettes and caps.
Specifications
Standards
Regulatory compliance stated for turbidity measurement (p.1 highlights; p.2 Technical Data, Regulatory Compliance), using an IR LED (860 nm) nephelometric 90° measuring principle
Tags and applications
- Parameter
- Turbidity
- Deployment
- LaboratoryField / Portable
- Brand
- Lovibond
Downloads
Common questions
What turbidity range does the Lovibond TB350 measure?
It measures 0.01 to 4,000 NTU with a full-range calibration from 0 to 4000 NTU. The display units are NTU, FNU, degrees, mg/L Kaolin and mg/L PSL.
What is the difference between the IR and white LED versions?
The IR versions use an 860 nm LED and the supplier lists ISO 7027 compliance. The white LED versions are EPA approved as an alternative to US EPA 180.1. Both measure at 90° with the same range.
How much sample does the TB350 need?
It needs 10 mL of sample per measurement.
How long does a reading take?
The response time is 7 seconds. Reading modes are Single, Signal Averaging and Fast-Settling.
Can I take the TB350 into the field?
The battery pack versions are portable and run on 3 NiMH AA cells. The instrument is IP 67 and the stated operating temperature is 0 to 50 °C. The power supply versions are sold as benchtop units.
Which calibration standards are supplied?
Every version includes T-CAL® standards of 5, 20, 800 and 4000 NTU with a test certificate. The white LED versions add a 2000 NTU T-CAL® standard.
Tell us what you need to measure.
Send the parameters and ranges you need, and we will confirm the right configuration and price. Your enquiry will name this product.