Lovibond · Meter (benchtop / portable)

TB211 IR

It measures turbidity in water samples from 0.01 to 1100 NTU, using infrared light scattered at 90° as described in EN ISO 7027. Use it as a portable turbidimeter that runs on a 9 V battery. It stores 125 data sets in its internal memory.

Measuring range0.01 - 1100 NTU
Accuracy± 2.5 % of reading, or ± 0.01 NTU, whichever is bigger (0.01 - 500 NTU); ± 5 % of reading (500 - 1000 NTU)
Measuring principlenephelometric (90° scattered light), EN ISO 7027
Light sourcetemperaturecompensated LED ( λ = 860 nm), infrared light
Internal storageinternal ring memory for 125 data sets
InterfaceMicro-USB
TB 211 IR turbidimeter in open carry case with calibration vials, light shield and USB cable

Highlights

01
Infrared 90° scattered-light measurement

A temperature-compensated 860 nm LED measures nephelometric scattered light at 90°, as provided in EN ISO 7027.

02
Autorange 0.01 to 1100 NTU

The detection limit is 0.01 NTU. Accuracy is ± 2.5 % of reading or ± 0.01 NTU, whichever is bigger, up to 500 NTU.

03
Stores 125 data sets

An internal ring memory holds 125 data sets, alongside a real-time clock and date. The interface is Micro-USB.

04
Portable unit on 9 V battery

It measures 110 x 55 x 190 mm, weighs 400 g as the basic unit and has splash-proof polycarbonate membrane operation.

Parameters

NTU

Turbidity

The range extends to 1100 NTU, but accuracy is only specified up to 1000 NTU, so readings above that carry no stated accuracy.

Measuring range0.01 - 1100 NTU
Accuracy± 2.5 % of reading, or ± 0.01 NTU, whichever is bigger, in the range of 0.01 - 500 NTU ; ± 5 % of reading in the range of 500 - 1000 NTU
Detection limit0.01 NTU
UnitsTE / F = NTU = FNU

Accuracy across the range

+0.01 … +500 NTU± 2.5 % of reading, or ± 0.01 NTU, whichever is bigger
+500 … +1000 NTU± 5 % of reading
+0.01 NTU+250.01+500.01+750+1000 NTU

Reagents & consumables

We keep these in stock for the TB211 IR. We quote part numbers per instrument on enquiry, checked against the current price list.

Standards & buffers

Calibration standards

Calibration standards, pH buffers and electrolyte solutions for calibrating the instrument.

Cells & glassware

Cuvettes & sample cells

Round cells, cuvettes and caps.

Specifications

Part Number266030
Measuring Range0,01-1100 NTU (Autorange)
Accuracy± 2.5 % of reading, or ± 0.01 NTU, whichever is bigger, in the range of 0.01 - 500 NTU ; ± 5 % of reading in the range of 500 - 1000 NTU
Measuring Principlenephelometric (90° scattered light)
Opticstemperaturecompensated LED ( λ = 860 nm) and photosensor amplifier in water proof sample chamber, infrared light
Measurement standardEN ISO 7027, infrared light at an angle of 90 °
Detection limit0.01 NTU
Displaybacklit LCD
InterfacesMicro-USB
Operationpolycarbonate membrane, splash proof
Auto – OFFYes
Internal Storageinternal ring memory for 125 data sets
Power Supply9 V-block
ClockReal Time Clock and Date
PortabilityPortable
Environmental Conditions5 - 40 °C at 30 - 90 % relative humidity (non condensing)
ComplianceCE
Dimensions110 x 55 x 190 mm
Weight400 g (Basic unit)
Delivery ScopeInstrument in carrying case; 4 turbidity standards (< 0,1, 20, 200 and 800 NTU); 9 V battery; 2 vials (ø 24 mm) with lids; USB cable 1.5 m; Warranty information; Certificate of Compliance; Instruction Manual

Standards

EN ISO 7027

Turbidity measured as scattered light at 90° with infrared light (860 nm), as provided in EN ISO 7027. Source: page 1 description and Technical Data table.

Tags and applications

Param TurbidityDeploy LaboratoryDeploy Field / PortableFormat Meter (benchtop / portable)Brand LovibondApp Drinking water treatment & distributionApp Swimming pools & water featuresApp Wastewater & trade effluent

Select any tag to see every other instrument that carries it. A tag with no link is one this line does not filter on.

Downloads

Common questions

What is the measuring range of the Lovibond TB211 IR?

The range is 0.01 to 1100 NTU with autorange. Accuracy is specified up to 1000 NTU, so ask us if you need to work between 1000 and 1100 NTU.

Which standard does the TB211 IR follow?

It uses the nephelometric method of EN ISO 7027. An 860 nm infrared LED measures scattered light at an angle of 90°.

What accuracy can I expect?

From 0.01 to 500 NTU it is ± 2.5 % of reading or ± 0.01 NTU, whichever is bigger. From 500 to 1000 NTU it is ± 5 % of reading.

Does it read in NTU or FNU?

The specification lists the units as TE/F = NTU = FNU. It gives no separate FNU range or accuracy.

What is supplied with the TB211 IR?

It comes in a carrying case with four turbidity standards (< 0.1, 20, 200 and 800 NTU) and a 9 V battery. Two 24 mm vials with lids, a 1.5 m USB cable, warranty information, a Certificate of Compliance and the instruction manual are also included.

How do I get stored readings off the instrument?

It keeps 125 data sets in an internal ring memory and has a Micro-USB interface. The data sheet does not describe transfer software or file format, so ask us before you plan a workflow.