Skip to content
Solution: TSS and CCTV monitoring with automatic intervention, for construction sites under Earth Control Measures.See how it works
Ebro · High-precision USB PDF data logger

EBI 310

The Ebro EBI 310 is a high-precision USB PDF data logger for recording temperature, and relative humidity on the EBI 310 TH, in places such as cold rooms, warehouses and transport. Plug it into a PC's USB port and it creates a PDF/A report of the readings, with no special software needed. Versions are available with an internal sensor, an external Pt 1000 probe, a humidity and temperature probe, or a two-channel adapter for exchangeable probes.

Measurement rangeVaries by versionsee Specifications
Accuracy±0.2 °C (-30 °C … +30 °C), internal sensor
Data memory120,000 readings
Sample rate1 s … 24 h
ReportPDF/A-1b, created on USB connection
Protection classIP65 · IP20varies by version
EBI 310 data logger

Highlights

01
±0.2 °C with a Pt 1000 sensor

The internal Pt 1000 sensor is accurate to ±0.2 °C from -30 to +30 °C, with 0.1 °C resolution.

02
PDF report on USB connection

Plug the logger into a PC's USB port and it writes a PDF/A-1b report of the readings. No special software is needed to read it.

03
From -200 °C to +400 °C with probes

The TE probe covers -200 to +250 °C. The TX takes up to two exchangeable probes, reaching +400 °C depending on type.

04
Five alarm ranges

Set up to five alarm ranges. A red and yellow LED shows the alarm status, and 120,000 readings fit in memory.

Parameters

T

Temperature

Measuring rangeEBI 310-30 °C … +75 °CEBI 310 TE-30 °C … +75 °C internal, -200 °C … +250 °C probeEBI 310 TH-30 °C … +75 °CEBI 310 TX-30 °C … +75 °C internal, -200 °C … +400 °C probes
AccuracyEBI 310±0.2 °C (-30 °C … +30 °C), ±0.5 °C (rest of range)EBI 310 TE probe±2.0 °C (-200 °C … -100 °C), ±1.0 °C (-100 °C … -20 °C), ±0.2 °C (-20 °C … +60 °C), ±0.5 °C (+60 °C … +250 °C)EBI 310 TH probe±0.5 °C (0 °C … +60 °C), ±0.8 °C (rest of range)EBI 310 TX probesdepend on type
Resolution0.1 °C
RH

Relative humidity (EBI 310 TH)

Measured by a capacitive sensor in the external probe.

Measuring range0 … 100 % rH
Accuracy±2 % rH (10 … 90 % rH at 25 °C), ±4 % rH (rest of range)
Resolution0.1 % rH

Configurations

All four have an internal Pt 1000 sensor and store 120,000 readings at intervals from 1 second to 24 hours.

1340-6331

EBI 310

Internal Pt 1000 sensor, -30 to +75 °C, ±0.2 °C from -30 to +30 °C. IP65.

Internal sensor; IP65
1340-6337

EBI 310 TE

Adds an external Pt 1000 probe on a 1 m PTFE cable, -200 to +250 °C. IP65.

External probe; IP65
1340-6336

EBI 310 TH

Adds an external probe for relative humidity, 0 to 100 %, and temperature. IP20.

Temperature and humidity; IP20
1340-6339

EBI 310 TX

A two-channel adapter for up to two exchangeable probes, up to +400 °C. Probes ordered separately. IP65.

Two probe channels; IP65

Specifications

Measurement rangeEBI 310-30 °C … +75 °CEBI 310 TE-30 °C … +75 °C internal, -200 °C … +250 °C probeEBI 310 TH-30 °C … +75 °C; 0 … 100 % rHEBI 310 TX-30 °C … +75 °C internal, -200 °C … +400 °C with probes, depending on type
Accuracy, temperatureEBI 310±0.2 °C (-30 °C … +30 °C), ±0.5 °C (rest of range)EBI 310 TEinternal as EBI 310; probe ±2.0 °C (-200 °C … -100 °C), ±1.0 °C (-100 °C … -20 °C), ±0.2 °C (-20 °C … +60 °C), ±0.5 °C (+60 °C … +250 °C)EBI 310 THinternal as EBI 310; probe ±0.5 °C (0 °C … +60 °C), ±0.8 °C (rest of range)EBI 310 TXinternal as EBI 310; probes depend on type
Accuracy, humidityEBI 310 TH±2 % rH (10 … 90 % rH at 25 °C), ±4 % rH (rest of range)
ResolutionEBI 310EBI 310 TEEBI 310 TX0.1 °CEBI 310 TH0.1 °C; 0.1 % rH
SensorEBI 310Pt 1000, internalEBI 310 TEPt 1000 probe, stainless steel, Ø 5 mm, 50 mm long, blunt, on a 1 m PTFE cable (waterproof, oil-proof, food-safe)EBI 310 THPt 1000 temperature and capacitive humidity probeEBI 310 TXtwo-channel adapter for up to two Pt 1000 probes, ordered separately
Data memory120,000 readings
Sample rate1 s … 24 h
Alarm5 ranges, shown by a red and yellow LED
ReportPDF/A-1b, created automatically when connected to a PC by USB
DisplayValue, min/max, alarm status
Storage temperature-40 °C … +85 °C
Housing materialPolycarbonate
DimensionsEBI 31080 x 33 x 14 mmEBI 310 TE91 x 33 x 14 mmEBI 310 TH129 x 33 x 14 mmEBI 310 TX111 x 33 x 14 mm
WeightEBI 310approx. 35 g
Protection classEBI 310EBI 310 TEEBI 310 TXIP65EBI 310 THIP20
BatteryLithium button cell CR 2450, 3 V; up to 2 years, depending on the application
CertificateFactory calibration certificate
Probes for the EBI 310 TXTPX 310-P1 (1341-6338): -200 °C … +200 °C, 45 mm x Ø 5 mm, 3 m PTFE cable TPX 310-P2 (1341-6339): -50 °C … +180 °C, 130 mm x Ø 3 mm, 3 m PTFE cable TPX 310-P3 (1341-6340): -50 °C … +180 °C, 130 mm x Ø 3 mm, 1 m PTFE cable TPX 310-P4 (1341-6341): +100 °C … +400 °C, 50 mm x Ø 1.5 mm, 3 m metal-wrapped cable, not waterproof TPX 310-P5 (1341-6342): -50 °C … +180 °C, 130 mm x Ø 3 mm, 5 m PTFE cable TPX 310-P6 (1341-6343): -50 °C … +180 °C, 130 mm x Ø 3 mm, 7.5 m PTFE cable TPX 310-P7 (1341-6344): -50 °C … +180 °C, 130 mm x Ø 3 mm, 10 m PTFE cable
Replacement probesEBI 310 TETPX 220 (1341-6332)EBI 310 THTPH 500 (1341-6337)EBI 310 TXTPX 310 adapter (1341-6335)
CE conformityComplies with CE guidelines 2014/30/EU
Temperature recorder standardEN 12830
EN 12830 suitabilityS (Storage), T (Transport)
EN 12830 locationC (Food storage and distribution systems)
EN 12830 accuracy classification1
Periodic inspectionYearly, as per EN 13486
Delivery scopeData logger; user manual

Standards

EN 12830

Temperature recorders for storage and transport of food. Suitability S (storage) and T (transport), location C (food storage and distribution systems), accuracy class 1. Yearly inspection per EN 13486.

Tags and applications

Select any tag to see every other instrument that carries it. A tag with no link is one this line does not filter on.

Downloads

Common questions

How accurate is the EBI 310?

The internal Pt 1000 sensor is accurate to ±0.2 °C from -30 °C to +30 °C, and ±0.5 °C over the rest of its range.

Which version measures humidity?

The EBI 310 TH. It measures 0 to 100 % rH, to ±2 % rH between 10 and 90 % rH at 25 °C.

Which version suits very low or very high temperatures?

The EBI 310 TE probe covers -200 °C to +250 °C. The EBI 310 TX reaches -200 °C to +400 °C, depending on the probes fitted.

Are probes included with the EBI 310 TX?

No. The TX takes up to two probes from the TPX 310-P1 to P7 range, ordered separately.

Is software needed to read it?

No. Connecting it to a PC by USB creates a PDF/A-1b report. Winlog.basic or Winlog.pro software can process the readings further.

How many readings can it store?

120,000 readings, at intervals from 1 second to 24 hours.

How long does the battery last?

Up to 2 years, depending on the application. It takes a CR 2450 lithium button cell.