Skip to content
Solution: TSS and CCTV monitoring with automatic intervention, for construction sites under Earth Control Measures.See how it works
Ebro · Validation data logger

EBI 12 steam steriliser

The Ebro EBI 12 steam steriliser loggers are validation data loggers for recording temperature during moist heat sterilisation, in places such as steam sterilisers and, with the T421 and T441, washer disinfectors. They measure with a PT1000 class A probe and transmit readings in real time, with an accuracy of ±0.05 °C between +120 °C and +140 °C. The versions differ in measurement range, probe length and channels, and five add EN 285 and EN 13060.

Measurement range-200 °C … +400 °Cvaries by version
Best accuracy± 0.05 °C (+120 °C … +140 °C)
Resolution0.01 °C
Data memory2x 50,160 · 4x 25,080 measurementsvaries by version
IP ratingIP68
polished ebro logger with two coiled rigid probes on a white plinth

Highlights

01
One logger, seven versions

Seven versions share a PT1000 class A sensor and 0.01 °C resolution, and differ in range, probe, memory and listed standards.

02
IP68 stainless steel housing

The 316 L stainless steel and PEEK housing measures 48 mm x 24 mm (Ø x H) and weighs approximately 90 g.

03
Interval range and real-time transmission

The measurement interval runs from 250 ms to 24 hours, and real-time data transmission is supported.

04
Factory calibration at +134 °C

Every version has a factory calibration certificate covering 0 °C, +60 °C and +134 °C, and lists ISO 17665 for moist heat sterilisation.

Parameters

T

Temperature

Accuracy is stated in bands across the range, and the tightest band, ±0.05 °C, applies between +120 °C and +140 °C.

Range-200 °C … +400 °C
Resolution0.01 °C
SensorPT1000, class A

Configurations

Every version has the same PT1000 class A sensor, 0.01 °C resolution, IP68 housing and 250 ms to 24 hour interval, and they differ in range, probe, memory and listed standards.

1340-6630

EBI 12-T421

Widest range, -200 °C … +400 °C, with 500 mm probe of 1.50 mm diameter. Lists food contact (EC) 1935/2004 and washer disinfector ISO 15883 standards.

1340-6629

EBI 12-T441

Measurement range -200 °C … +200 °C, with ± 1.5 °C accuracy from -200 °C to -90 °C. Otherwise matches the base version.

1340-6640

EBI 12-T470

Range -20 °C … +150 °C, four measuring channels, 1.20 mm x 600 mm probe. Adds EN 285 and EN 13060; no food contact or washer disinfector standard.

1340-6644

EBI 12-T471

Range -20 °C … +150 °C with 1.20 mm diameter, 1,200 mm long probe. Adds EN 285 and EN 13060; no food contact or washer disinfector standard.

1340-6634

EBI 12-T490

Range -20 °C … +150 °C with 1.20 mm diameter, 600 mm long probe. Adds EN 285 and EN 13060; no food contact or washer disinfector standard.

1340-6645

EBI 12-T671

Range -20 °C … +150 °C, four measuring channels, data memory 4x 25,080 measurements, 1.20 mm probe. Adds EN 285 and EN 13060.

1340-6635

EBI 12-T690

Range -20 °C … +150 °C, data memory 4x 25,080 measurements, 1.20 mm x 600 mm probe. Adds EN 285 and EN 13060.

Specifications

Operating temperatureEBI 12-T421EBI 12-T441-55 °C … +150 °CEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T690-20 °C … +150 °C
Measurement rangeEBI 12-T421-200 °C … +400 °CEBI 12-T441-200 °C … +200 °CEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T690-20 °C … +150 °C
AccuracyEBI 12-T421± 0.5 °C (-90 °C … -40 °C) ± 0.2 °C (-40 °C … 0 °C) ± 0.1 °C (0 °C … +120 °C) ± 0.05 °C (+120 °C … +140 °C) ± 0.1 °C (+140 °C … +150 °C) ± 0.5 °C (+150 °C … +250 °C) ± 0.8 °C (+250 °C … +400 °C)EBI 12-T441± 1.5 °C (-200 °C … -90 °C) ± 0.5 °C (-90 °C … -40 °C) ± 0.2 °C (-40 °C … 0 °C) ± 0.1 °C (0 °C … +120 °C) ± 0.05 °C (+120 °C … +140 °C) ± 0.1 °C (+140 °C … +150 °C) ± 0.5 °C (+150 °C … +200 °C)EBI 12-T470EBI 12-T671± 0.2 °C (-20 °C … 0 °C); ± 0.1 °C (0 °C … +120 °C); ± 0.05 °C (+120 °C … +140 °C); ± 0.1 °C (+140 °C … +150 °C)EBI 12-T471EBI 12-T490EBI 12-T690± 0.2 °C (-20 °C … 0 °C) ± 0.1 °C (0 °C … +120 °C) ± 0.05 °C (+120 °C … +140 °C) ± 0.1 °C (+140 °C … +150 °C)
Resolution0.01 °C
Data memoryEBI 12-T421EBI 12-T441EBI 12-T470EBI 12-T471EBI 12-T4902x 50,160 measurementsEBI 12-T671EBI 12-T6904x 25,080 measurements
SensorPT1000, class A
Measurement interval250 ms … 24 hours
Real-time data transmissionYes
Storage temperature-20 °C … +70 °C
Measurement modesEBI 12-T421EBI 12-T441EBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671Endless measurement; Measure upon start time; Start immediately until end of memory; Start- / Stop measurementEBI 12-T690endless measurement; measurement from the start time; Measure software until memory is full; Start- / Stop measurement
BatteryLithium, 3.6 V, User-replaceable
Dimensions (Ø x H)48 mm x 24 mm
WeightApp. 90 g
Housing material316 L stainless steel / PEEK
IP ratingIP68
Factory calibration certificateEBI 12-T421EBI 12-T441-20 °C, 0 °C, +60 °C and +134 °CEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T6900 °C, +60 °C and +134 °C
Food contact materials regulationEBI 12-T421EBI 12-T441(EC) 1935/2004 – Materials and articles which come into contact with foodEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T690Not stated
Sterilization standardISO 17665 – Sterilization of health care products — Moist heat
Washer disinfector standardEBI 12-T421EBI 12-T441ISO 15883 – Washer DisinfectorsEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T690Not stated
Radio equipment directive2014/53/EU – Radio Equipment Directive RED
Probe diameterEBI 12-T421EBI 12-T4411.50 mmEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T6901.20 mm
Probe lengthEBI 12-T421EBI 12-T441500 mmEBI 12-T470EBI 12-T490EBI 12-T690600 mmEBI 12-T471EBI 12-T6711,200 mm
Steam sterilizer standardEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T690EN 285 – Sterilization, steam sterilizers, medical equipmentEBI 12-T421EBI 12-T441Not stated
Small steam sterilizer standardEBI 12-T470EBI 12-T471EBI 12-T490EBI 12-T671EBI 12-T690EN 13060 – small steam sterilizersEBI 12-T421EBI 12-T441Not stated

Standards

ISO 17665

Sterilization of health care products — Moist heat

ISO 15883

Washer Disinfectors

Tags and applications

Select any tag to see every other instrument that carries it. A tag with no link is one this line does not filter on.

Downloads

Common questions

What temperature range does the EBI 12 steam steriliser validation logger cover?

The EBI 12-T421 measures -200 °C to +400 °C and the EBI 12-T441 measures -200 °C to +200 °C. The EBI 12-T470, T471, T490, T671 and T690 measure -20 °C to +150 °C.

How accurate is it around 134 °C?

Accuracy is ±0.05 °C between +120 °C and +140 °C on every version. Resolution is 0.01 °C.

How many measurements can it store?

The EBI 12-T421, T441, T470, T471 and T490 store 2x 50,160 measurements. The EBI 12-T671 and T690 store 4x 25,080 measurements.

Can the battery be replaced by the user?

Yes. It uses a 3.6 V lithium battery that the user can replace.

Which standards are listed for the EBI 12?

All versions list ISO 17665 for moist heat sterilisation. The EBI 12-T421 and T441 also list ISO 15883 for washer disinfectors and (EC) 1935/2004 for food contact materials. The EBI 12-T470, T471, T490, T671 and T690 list EN 285 and EN 13060.

What is the IP rating and housing material?

The logger is rated IP68. The housing is 316 L stainless steel and PEEK.

Tell us what you need to measure.