XD 7000 / XD 7500
The Lovibond XD 7000 / XD 7500 is a benchtop spectrophotometer for water analysis in places such as drinking water treatment, swimming pools, cooling towers and wastewater plants. It runs more than 150 pre-programmed methods, including chlorine, COD, ammonia and turbidity, and also records absorbance, % transmission, kinetics and spectra. The XD 7000 covers the visible range, and the XD 7500 adds the UV range and the SAC 254 nm method.
Highlights
The XD 7000 covers 320 - 1100 nm and the XD 7500 covers 190 - 1100 nm, with 1 nm wavelength resolution.
Accuracy is 0.003 Abs below 0.6 Abs and 0.5 % from 0.6 to 2.0 Abs. Linearity is under 1 % up to 2.0 Abs.
Methods cover chlorine, COD, ammonia, nitrate, phosphate, metals and turbidity. The memory also holds up to 100 user programmes.
Recognises round (13, 16, 24 mm) and rectangular (10, 20, 50 mm) cuvettes. An internal barcode reader identifies the test.
Methods and ranges
The table shows a selection of the more than 150 methods. The methods shown are photometric water analyses, mostly with tablet or powder-pack reagents, and the full list is in the datasheet download.
| Parameter | Range | Method | Reagent |
|---|---|---|---|
| Alkalinity-m T | 5 - 200 mg/L CaCO3 | Acid / Indicator | Tablet |
| Ammonia LR TT | 0.02 - 2.5 mg/L N | Salicylate | |
| Chlorine dioxide T | 0.02 - 11 mg/L ClO2 | DPD / Glycine | Tablet |
| Chlorine T | 0.01 - 6.0 mg/L Cl2 | DPD | Tablet |
| COD LR TT | 3 - 150 mg/L COD | Dichromate / H2SO4 | |
| COD MR TT | 20 - 1500 mg/L COD | Dichromate / H2SO4 | |
| Copper T | 0.05 - 5 mg/L Cu | Biquinoline | Tablet |
| Iron T | 0.02 - 1 mg/L Fe | Ferrozine / Thioglycolate | Tablet |
| Nitrate LR TT | 0.5 - 14 mg/L N | 2,6-Dimethylphenole | |
| pH-value T | 6.5 - 8.4 pH | Phenol Red | Tablet |
| Phosphate PP | 0.02 - 0.815 mg/L P | Phosphomolybdenum Blue | Powder pack |
| SAC 254 nm (344) | 0.25 - 50 m-1 | Direct Reading | |
| Suspended solids 50 | 10 - 750 mg/L TSS | Turbidity / Attenuated Radiation Method | |
| Turbidity 50 | 5 - 500 FAU | Attenuated Radiation Method | |
| Zinc T | 0.02 - 1 mg/L Zn | Zincon | Tablet |
Accuracy across the range
Accuracy over each temperature range. The bar shows where the range sits on the scale.
Configurations
Both versions are benchtop grid-monochromator spectrophotometers with the same display, interfaces and delivery scope, and differ in lamp, wavelength range and stray light.
XD 7000 (VIS)
Visible-range model with tungsten halogen lamp, 320 - 1100 nm wavelength range and stray light below 0.1 % Transmission at 340 and 408 nm.
XD 7500 (UV-VIS)
Xenon flash lamp (500 million flashes possible), 190 - 1100 nm range, stray light below 0.05 % Transmission; adds SAC 254 nm method.
Reagents & consumables
These are the reagents and consumables this instrument uses. We confirm the part numbers for your instrument when we quote.
Liquid reagents, Powder packs, Tablet reagents, Tube tests
Tablets, pre-measured powder packs, liquid reagents and sealed tube tests for the methods already programmed into the instrument.
Calibration standards
Calibration standards, pH buffers and electrolyte solutions for calibrating the instrument.
Cuvettes & sample cells
Round cells, cuvettes and caps.
Specifications
Tags and applications
Downloads
Common questions
What wavelength range does the XD 7000 / XD 7500 cover?
The XD 7000 (VIS) covers 320 - 1100 nm. The XD 7500 (UV-VIS) covers 190 - 1100 nm. Wavelength accuracy is ± 1 nm on all Holmium peaks.
What is the difference between the XD 7000 and the XD 7500?
The XD 7000 has a tungsten halogen lamp and a 320 - 1100 nm range. The XD 7500 has a xenon flash lamp and a 190 - 1100 nm range. Stray light is below 0.1 % Transmission on the XD 7000 and below 0.05 % Transmission on the XD 7500, both at 340 and 408 nm. The XD 7500 also adds the SAC 254 nm method.
How many pre-programmed methods does it have?
More than 150. The method table on this page shows a selection of them. The full list is in the datasheet download.
Can it export results to a LIMS?
Yes. It supports LIMS transfer as ASCII and .csv files. Interfaces are Ethernet, USB B, USB A for external memory, keypad, barcode scanner and a PCL compatible printer.
Which cuvettes can it use?
It recognises cuvettes automatically. Round cuvettes of 13, 16 and 24 mm are supported. Rectangular cuvettes of 10, 20 and 50 mm are supported.
What is the photometric accuracy?
Photometric accuracy is 0.003 Abs below 0.6 Abs. From 0.6 to 2.0 Abs it is 0.5 %. The photometric range is -3.3 to +3.3 Abs.
Tell us what you need to measure.
Send the parameters and ranges you need, and we will confirm the right configuration and price. Your enquiry will name this product.