inoLab Cond 7110 / Cond 7310
The WTW inoLab Cond 7110 and Cond 7310 are benchtop conductivity meters for laboratories testing water and solutions, such as process water and food production samples. They work with two and four electrode cells, standard and special, and suit pure water measurements according to pharmacopoeia as well as highly concentrated solutions. They also determine salinity, resistivity and TDS, and the Cond 7310 adds data memory and GLP/AQA documentation through USB or an optional built-in printer.
Highlights
Works with standard and special two and four electrode cells, for pure water through to highly concentrated solutions.
Conductivity readings are supplemented by salinity, specific resistance and TDS, with temperature measured from -5.0 to 105.0 °C.
The meter is suitable for pure water measurements according to pharmacopoeia.
AutoRead recognises a stable value, and an adjustable calibration timer reminds the user when to calibrate.
Parameters
Conductivity
The lowest conductivity ranges depend on the cell constant of the cell connected, 0.01 cm-1 or 0.1 cm-1.
Resistivity
Salinity
Total dissolved solids
Temperature
Configurations
Both models measure conductivity, resistivity, salinity, TDS and temperature with two- and four-electrode cells.
Cond 7110
LCD display, adjustable cell constants up to 2.500 cm-1 and TDS up to 1999 mg/l.
Cond 7310
Backlit graphic display, data memory, USB, GLP/AQA support, cell constants up to 25.000 cm-1 and TDS to 199.9 g/l; printer optional.
Consumables
These are the consumables this instrument uses. We confirm the part numbers for your instrument when we quote.
Calibration standards
Calibration standards, pH buffers and electrolyte solutions for calibrating the instrument.
Electrodes & probes
Electrodes, probes and replacement sensors, matched to the instrument.
Specifications
Tags and applications
- Parameter
- Conductivity / TDS
- Deployment
- Laboratory
- Brand
- WTW
Downloads
Common questions
What conductivity range do the inoLab Cond 7110 and Cond 7310 measure?
It measures 0.00 to 1000 mS/cm at ±0.5 % of value. Low-conductivity ranges are 0.000 to 1.999 µS/cm with a K of 0.01 cm-1 and 0.00 to 19.99 µS/cm with a K of 0.1 cm-1.
What is the difference between the Cond 7110 and the Cond 7310?
The Cond 7310 has a backlit black and white graphic display, where the Cond 7110 has an LCD. The Cond 7310 also has a Mini USB-B interface, memory, GLP/AQA support, a printer option and 10 calibration records against 1 on the Cond 7110. Its TDS range extends to 199.9 g/l and its adjustable cell constants to 25.000 cm-1.
Can the Cond 7310 store readings and connect to a PC?
Yes. The Cond 7310 holds 500 manual or 5000 automatic memory entries and has a Mini USB-B interface. The Cond 7110 has neither memory nor an interface.
Which cell constants can I use?
The fixed cell constant is 0.01 cm-1. With calibration it accepts 0.450 to 0.500 cm-1 and 0.800 to 0.880 cm-1. Adjustable ranges are 0.090 to 0.110 cm-1 and 0.250 to 2.500 cm-1 on the Cond 7110, and 0.090 to 0.110 cm-1 and 0.250 to 25.000 cm-1 on the Cond 7310.
How does temperature compensation work?
Compensation can be automatic or switched off, with a reference temperature of 20 °C or 25 °C. The setting range is none, nlF or 0.000 to 3.000 %/K on the Cond 7110, and none, nlF or 0.000 to 10.000 %/K on the Cond 7310.
Does it run on batteries or mains power?
It runs on a battery or on the universal power supply.
Tell us what you need to measure.
Send the parameters and ranges you need, and we will confirm the right configuration and price. Your enquiry will name this product.